Spectrophotometric Determination Of Iron


Spectrophotometric Determination Of Iron
INTRODUCTION
The direct-reading, single-beam spectrophotometer. Is most suitable for fast operation, and because of their single beam design they are relatively inexpensive and require a minimum of maintenance. They are particularly well suited for spectrophotometric determinations of a single component at a single wavelength where only moderate accuracy ( ±1 to 3%T) is required.
The limitations of the single-beam instrument become apparent when an absorption spectrum over a wavelength range is required. The response of the phototube, the emissivity of the light source, and the intensity of the light diffracted by the grating are all a function of wavelength.
Accordingly, in order to obtain the absorption spectrum of a compound, the instrument must be recalibrated each time the wavelength setting is changed.
HANDLING OF CUVETTES
The handling of the cuvettes is extremely important. Often two cuvettes are used simultaneously,
one for the "blank" solution and one for the samples to be measured. Yet any variation in the cuvette (such as achange in the cuvette width or curvature of the glass, stains, smudges, or scratches) will cause varying results. Thus, it is essential to follow several rules in dealing with cuvettes:
1. Do not handle the lower portion of a cuvette through which the light beam will pass.
2. Always rinse the cuvette with several portions of the solution before taking a measurement.
3. Wipe off any liquid drops or smudges on the lower half of the cuvette with a clean Kimwipe
(or other lens paper) before placing the the cuvette in the instrument. Never wipe the cuvette
with paper towels or handkerchiefs. Inspect the cuvette to ensure that nor bubbles are
clinging to the inside walls.
4. When inserting a cuvette into the sample holder:
a. To avoid any posible scratching of the cuvette in the optical path, insert the cuvette
withthe index line facing toward the front of the instrument.
b. After the cuvette is seated, line up the index lines exactly.
5. When using two cuvettes simultaneously, use one of the cuvettes always for the blank
solution and the other for the various samples being measured. Mark the cuvette ( on the
upper portion) accordingly and do not interchange the cuvettes during the remainder of the experiment.
6. Do not use a test tube brush to clean the cuvettes after the experiment is complete.
REAGENTS
1,10-phenanthroline (0.1 g of 1,10-phenanthroline monohydrate in 100 mL of distilled
water, warming to effect solution if necessary).
Hydroxylamine hydrochloride (10 g in 100 mL of distilled water).
Sodium acetate (10 g in 100 mL of distilled water).
Ferrous ammonium sulfate hexahydrate.
PART A -- Determination of Spectral Response of a Single-Beam Spectrophotometer
The purpose of this section is to familiarize you with the single beam spectro photometer operation and to provide a justification for why you must reset your instrument response each time a newwavelength has been chosen
PROCEDURE
warm-up the instrument before analysis ,read instrument manual for warm up time, latest instruments required less time for warm up before use. First, adjust the
instrument, photocell dark, with the zero control. Next, insert a cuvette filled with water into the sample holder. Take blank reading absorbance or .transmittance. Without changing either the dark current control or the 100%T control, vary the
wavelength from 340 to 900 nm in 4 nm intervals and record the instrument response. Plot the instrument response vs. wavelength.
PART B -- Determination of Iron with 1,10-phenanthroline
INTRODUCTION
The reaction between ferrous ion and 1,10-phenanthroline to form a red complex serves as a sensitive method for determining iron.
Fe2+ + 3 1,10-phenanthroline ® Fe(phen)3 2+
1,10 – phenanthrolene (Phen)
The molar absorptivity of the complex, [(C12H8N2)3Fe]2+, is 11,100 at 508 nm. The intensity of the color is independent of pH in the range 2 to 9. The complex is very stable and the color intensity does not change appreciably over long periods of time. Beer's law is obeyed.
The iron must be in the ferrous state, and hence a reducing agent is added before the color is developed. Hydroxylamine, as its hydrochloride, can be used to reduce any ferric ion that is present:
2 Fe3+ + 2 NH2OH + 2 OH- ® 2 Fe2+ + N2 + 4 H2O
The pH is adjusted to a value between 6 and 9 by addition of ammonia or sodium acetate. PROCEDURE
1. Weigh accurately about 0.07 g of pure ferrous ammonium sulfate hexahydrate, dissolve it in water, and transfer the solution to a 1-liter volumetric flask. Add 2.5 mL of concentrated sulfuric acid and dilute the solution to the mark. Calculate the concentration of the solution in mg of iron per mL. (Remember, your solution was prepared using Fe(NH4)2(SO4)2·6H2O).
2. Prepare the unknown sample as follows. Add about 0.12 g of the solid unknown and
approximately 0.25 mL concentrated sulfuric acid into a 100 mL volumetric flask and dilute to the mark. Now transfer a 1 mL aliquot of this solution to another 100 mL volumetric flask do not dilute yet. This will be referred to as the "prepared unknown".
3. Into five 100 mL volumetric flasks, pipet (volumetrically) 1, 5, 10, 25, and 50 mL portions of the standard iron solution. Put 50 mL of distilled water into another flask to serve as the blank. To each flask, including the "prepared unknown", add 1 mL of the hydroxylamine solution, 10 mL of the 1,10 phenanthroline solution and 8 mL of the sodium acetate solution. Then dilute all the solutions to the 100 mL marks and allow them to stand for 10 minutes with occasional shaking of the flasks.
4. Using the blank as a reference and any one of the iron solutions prepared above, measure the
absorbance at different wavelengths in the interval from 340 to 900 nm . (Note that it is
necessary to re-adjust the 100%T setting whenever the wavelength is changed). Take reading  about 10 nm apart except in the region of maximum absorbance where intervals of 5 nm should be used. Plot the absorbance vs. wavelength and connect the points to from a smooth curve. Select the proper wavelength to use for the determination of iron with 1,10- phenanthroline.
5. Also, calculate the molar absorption coefficient, e, at the maximum absorption on the
absorption curve (assume b = 1 cm).
6. Measure the absorbance of each of the standard solutions and the unknown at the selected
wavelength. Plot the absorbance vs. the concentration of the standards. Note whether Beer's law is obeyed. Using the absorbance of the unknown solution calculate the % (w/w) iron in your original solid sample, remember to correct for dilutions.

2 comments:

Jack Smith said...

Spectrophotometry is one of the most important practice. spectrometer makes measuring intensity of light easy and information you shared is good.
Spectrophotometer cuvettes

Jack Smith said...

Good post....thanks for sharing.. very useful for me i will bookmark this for my future needs. Thanks.
Link - http://sharetforest.blogspot.in/2014/03/preview-cuvette-responsive-
multipurpose.html

m ginger